by admin / Thursday, 05 January 2023 / Published in blog

At present, technology transfer in general, and transfer analytical method, in particular, is a problems of great concern for factories or laboratories.

The transfer of analytical methods to improve product quality as well as bring domestic technology to ensure fast and sustainable development.

Before an analytical method is transferred, it should be properly developed and validated. The implementation of the transfer process, the analytical procedure, also known as the transfer of the analytical method, must be fully documented with the attached conditions. This article will discuss the key components required of the method transfer process between the receiving and transfer laboratories. Figure 1 describes analytical method transfer process.

Figure1: Analytical Method Transfer Professor

Each analytical method transfer between two laboratories involves two main parties: namely, the transferring laboratory (TL) and the receiving laboratory (RL). The transferring laboratory (TL) is the source or laboratory of origin for the analytical procedure, the receiving laboratory (RL) is the transfer laboratory’s recipient

Before transferring a test method, the transfer laboratory site should verify that the test method complies with those described in the marketing authorization or the relevant technical dossier.

The original validation of the test method(s) should be reviewed to ensure compliance with current requirements. Technical and operational risk assessments should be performed and documented to identify any supplementary validation that should be performed and from that to plan analytical method transfer.

An analytical method transfer plan that assesses time and resources, is recommended for the transfer of two or more methods. In some cases, when analytical method transfer is part of the transfer process, the analytical transfer method plan is included in the overall transfer plan.

When determining the transfer strategy, the following should be assessed: (1) Experience and knowledge of the receiving unit, (2) Degree of familiarity of the receiving unit with the methodology or technology used, (3) Specification of the product, and (4) Complexity of the analytical method. The analytical method transfer procedures may be different depending on the results of these assessments.

The nature of the method, the experimental set-up, and the data analysis must be tailored to the respective situation to ensure that regulatory requirements are met. A method transfer can be greatly simplified if, for example, the receiving laboratory (RL) already has experience with implementing a similar method, different strengths, or administration forms of the same product.

The following (Table 1) describes the different types of method transfer categories based on USP 1224:

Table 1: Different types of method transfer strategy with examples

Type of strategyDesignExamples
Comparative studiesThe same sample is tested in both the TL and the RL and the results are then comparedLC/GC assay and related substances
Other methods: e.g. water, residual solvents, ions, particle size distribution
Co-validationFrom the outset, the RL is included in the validation of the method being transferred.LC/GC assay and related substances  
RevalidationRL must performed revalidation or partial revalidation of the analytical method that is transferredMicrobiological testingOther critical threshold tests
Transfer waiverAnalytical procedures at RL are the same as TLCan be considered for all method transfers but requires scientific justification since testing is not performed

The analytical method transfer plan should include but is not limited to the following:

  1. Objectives
  2. The scope and responsibilities
  3. References
  4. Analytical Method Training
  5. Analytical Method Transfer Strategy
  6. Analytical Method Validation Strategy (as required)
  7. Qualification of Reference Standards
  8. Assessment of Critical Reagents
  9. Analytical Method Validation History
  10.  Sample Testing Matrix

An analytical method transfer protocol is prepared detailed and approved between the transfer laboratory (TL) and the receiving laboratory (RL) which should include, but is not limited to, the following:

  1. Purpose
  2. Scope
  3. Materials and Equipment
  4. Experimental design
  5. References
  6. Appendices

The receiving laboratory performs the transfer according to the approved analytical method transfer protocol. Data review, and assessment of results should be performed by the transfer laboratory (TL) and the receiving laboratory with the following issues to be considered:

  • Is the data being produced in the receiving laboratory considered fit for purpose?
  • Is the analytical method operating reliably in the receiving laboratory?
  • Are there any improvements to the analytical method to consider?

And in addition, deviations from the protocol should be investigated before the closure of the transfer process if applicable.

Acceptance criteria are also an important part of the successful transfer of analytical methods and must therefore be specified. The acceptance criteria should be based on the current validation study of the methodology and current requirements.

Thus, analytical method transfer is the most important part of a product transfer process, a decision on the success of the transfer process. The results of the transfer of analytical methods depend on many factors, so it requires careful preparation before the transfer to reach the goal of the successful transfer of analytical methods.

Documents references:

  1. USP <1224> Transfer of Analytical Procedures.
  2. ISPE Baseline Guide: Technology Transfer (Third Edition)
Tagged under: Laboratory, tvtpi

What you can read next

Mean Kinetic Temperature – MKT
The approval process of medicines in Europe
Tracebility Standards Process In Quality Control Department Of Pharmaceutical Manufacturing

Search

Recent Posts

  • ORIOLED HUB EMPOWERS THE FUTURE FOR STUDENTS OF HO CHI MINH CITY UNIVERSITY OF MEDICINE AND PHARMACY
    In event
    15 May, 2025
  • ORIOLED HUB INSPECTS RAW PHARMACEUTICAL SUPPLIERS IN CHINA – STRENGTHENING THE QUALITY SUPPLY CHAIN FOR THE REIWAPHARM FACTORY PROJECT
    In event
    15 May, 2025
  • TPI TELEVISION COLLABORATED WITH THONG NHAT HOSPITAL TO SUCCESSFULLY ORGANIZE THE STROKE CONFERENCE WITH OVER 300 MEDICAL STAFF IN ATTENDANCE.
    In event
    28 March, 2025
  • ‘Bringing the Moon to the Village’ with TV TPI: Mid-Autumn Festival for Mountainous Children
    In event
    18 November, 2024

TV TPI provides full package services from idea to market product that suits your business needs.

  • Home
  • About us
  • Product
  • Blog
  • Event
  • Recruitment

business@tvtpi.com.vn

+84 2838 663399

TV TPI: 72 Binh Gia Street, Ward 13, Tan Binh District, Ho Chi Minh City, Vietnam

OHI: No. 61A - 63A, Vo Van Tan Street, Vo Thi Sau Ward, District 3, Ho Chi Minh City, Vietnam

DMCA.com Protection Status

© Copyright 2024 TVTPI. All rights reserved.

TOP